The 12707 is a sample box used together with an electrometer to measure the volume resistivity of liquid. It requires only 0.8 cc of sample for measurement.
The electrodes can be removed and cleaned easily.
Operating environment: Temperature 0 ℃ to +80 ℃, relative humidity 80 % or less without codensation
Storage environment: Temperature -25 ℃ to + 80 ℃, relative humidity 90 % or less without codensation
Insulation resistance (temperature:23±5 ℃, relative humidity: 70% or less, 60 VDC):
Main electrode - Guard electrode 1×1013Ω or higher
Main electrode - Counter electrode 1×1013Ω or higher
Guard electrode - Counter electrode 1×109Ω or higher
Main electrode - Chassis 1×1013Ω or higher
Guard electrode - Chassis 1×109Ω or higher
Counter electrode - Chassis 1×109Ω or higher
Dielectric strength:
Counter electrode - Chassis 60 VDC (continuous)
Guard electrode - Chassis 60 VDC (continuous)
Required sample amount: 0.8 cc
Connector: INPUT terminal triaxial connector
V terminal biding post
LID terminal BNC connector
Electrode coefficient: 147.65
Dimensions: Approx. 150 (W)×80 (H)×150 (D) mm
Mass: Approx. 1.5 kg
Name | Part Number | Quantity |
---|---|---|
Connection cable | DCB-PP1801X01 | 3 |
Name | Part Number |
---|---|
Electrode for 12707 | CC019001 |