Scanner/Extender
3100/3101
Two switch cards can be mounted on the mainframe and three cards on each extender. Up to five externders can be connected by the scan buses. The distance between each device is up to 10 meter, and the total length is 20 m at a maximum. The extenders can be mounted according to the necessary number of measurement channels, allowing you to build cost-effective system configurations.
The 3100 adopts a large LCD screenm and displays the OPEN/CLOSE status of each modular card. The OPEN/CLOSE setting is available by local operation using the front panel.
The local function to set the switching conditions on the front panel is suitable for small-scale systems comblined with digital voltmeters, digital recorders or other devices.
On the other hand, those switching conditions also can be set from the GPIB or USB interface, allowing you to build up automatic measuring systems flexibly.
In addition, not only sequential operation using control parameters, but also direct channel access for arbitrary channels and direct program access for arbitrary program steps are available. This direct access function is operationable by local or remote control, and random scan operation of 100 steps or more can be easily performed by using the GPIB or USB interface.
The 3100 is equipped with the digital I/O or DIO functin. It adopts an open collector output and can output sink current up to 150 mA, so that it can be used to control lamps and large-sized relays.
It suppors the input level from the TTL level up to 28 V. The digial I/O function is aslo convinient for data transfer with devices that have no GPIB interface.
The 3100 is provided with seven switch cards of three types, and functions as multiple scanners by itself in combination with these switch cards.
Thus, it offers flexible solutions as data logger for multiple points and switching device of an automatic test system such as for semiconductors, electronic components and electronic circuits.
The 3100 has 2 slots, in which maximum 160 channels in total are available when using two 31001F by 1 wire.
By connecting five 3101 extenders, the configuration can be expanded up to 17 slots, that is, maximum 1,360 channels (1-wire 31001F).
This figure is a configuration example of an I-V characteristic measurement system for exposure test on forty solar panels.
Four 3101 extenders are connected to the 3100 scanner.
The forty solar panels are connected in 4-wire connection by using four large-current cards of the 31001K to eliminate errors in the cables, and then the I-V characteristics are measured in turn.
Prepare two sets of the 31005B and the 31009M in combination and connect them as multiplexor to the temperature sensors next to the solar panels. By using the measured temperature, the I-V measured values can be corrected.
This is a configuration example of integrated test systems for analog ICs. The crosspoints of the matirix cards can be opened or closed as desired. Also, any input or output of the X direction can be onnected to any input or output of the Y direction. Therefore, by connecting measuring and sourcing devices to the Y directon and the pins of an IC to the X direction, the inputs and outputs of those devices can be connected to any pin of the IC. You can use cards in combination according to the number of devices and the number of IC pins.
This is an configuration example of automatic temperature coefficient measurement systems having forty resitors. The 4-wire measurement method is used for low-resistnance measurement. The temperautre of the constant temperature chamber is controlled by the DV voltage source and is monitored by the digital multimeter having temperature measurenent function. Thus, the resistance values is measured at each temperature and is stored by the controller. Then, the temperature coefficient of each resistor is calculated from the difference in measurement temperature and reistance value variation, and is printed by the printer.
The digital I/O function of the 3100 has 16-bit interactive input and output lines, and the input and output can be set by 8 bits. Thus, eight lines can be used for controlling and other eight lines are for status monitoring. Otherwise, all the sixteen bits can be set as input or output.
Model |
Feature |
|
Switch card | 31001F | High-speed multiplexer card |
31001K | Large-current multiplexer card | |
31001G | High-voltage multiplexer card | |
31001H | 500 V withstand voltage multiplexer card | |
31001J | Micro-current multiplexer card | |
31002K | Large-current actuator card | |
31003B | General-purpose matrix card | |
31005B | Multiplexer/actuator combo card | |
Terminal card | 31009B | General-purpose terminal card |
31009M | Terminal card for 31005B multiplexer card |
This card is the 4-wire 20 channel type using semiconductor relays. It can be uses as 2-wire 40 channel or 1-wire 80 channel by setting.
Specifications
Number of channels |
20 ch/4-wire 40 ch/2-wire 80 ch/1-wire |
Channel switch | Semiconductor relay |
Peak voltage |
140 V (AC) 200 V (DC) |
Current |
(100 - operating temperature) mA |
Insulating resistance (between terminals)* |
1010 Ω or higher |
Thermal electromotive force | 10 μV or less (typical value) |
Number of required slots | 1 |
Applicable switch card | 31009B |
*At 23±5 ℃ and 60 % RH or less
Contact configuration
Current conducts between each channel and COM (A, B, C, D) by setting ON (closing) one of CH0 to CH9.
The input and output terminl section and the switch section are integrated. This card is a 4-wire 10 channel type. The terminals A and B are for large-current measurement and the terminals C and D are for small-current measurement.
Specifications
Number of channels |
10 ch/4-wire (2 wires for large current and and other 2 for small current) |
|
Channel switch |
Mechanical relay |
|
Peak voltage |
220 V (AC) 220 V (DC) |
|
Current | 10 A | 2 A |
Insulating resistance (between terminals)* |
1010 Ω or higher |
|
Thermal electromotive force | 50 μV or less | |
Number of required slots | 2 | |
Applicable switch card | Integrated type |
*At 23±5 ℃ and 60 % RH or less
Contact configuration
Current conducts between each channel and COM (A, B, C, D) by setting ON (closing) one of CH0 to CH9.
The card is a 1-wire 10 channel type, and the maximum contact voltage is 1000 V. Each channel has a terminal with a short-circuit current limiting resistor. It can be expanded up to 170 channels.
Specifications
Number of channels | 10 ch/1-wire |
Channel switch | Dry reed relay |
Peak voltage |
1000 V (AC) |
Current |
0.1 A (DC) 17 mA when using terminal R |
Insulating resistance (between terminals)* |
1010 Ω or higher |
Thermal electromotive force | 100 μV or less |
Number of required slots | 1 |
Applicable switch card | Integrated type |
*At 23±5 ℃ and 60 % RH or less
Contact configuration
Current conducts between each channel and COM by setting ON (closing) one of CH0 to CH9.
This multiplexer card has a maximum contact voltage of 500 V and a thermal EMF of less than 1 μV.
It occupies one slot with 4 wires x 10 channels or 2 wires x 20 channels, so that a space-saving system can be built with 40 channels by two cards on the 3100, or 100 channels by five cards on the 3100+3101.
Specifications
Number of channels |
10 ch/4-wire 20 ch/2-wire |
|
Channel switch |
Mechanical relay |
|
Peak voltage |
500 V (DC) 250 V (AC) |
|
Current |
1 A (DC30 V or less, AC62.5 V or less) |
|
Insulating resistance (between terminals)* |
1010 Ω or higher 108 Ω or higher |
|
Thermal electromotive force | 1 μV or less | |
Number of required slots | 1 | |
Applicable switch card | Integrated type |
*Uppwer row: At 23±5 ℃ and 60 % RH or less
Lower row: At 0 to 40 ℃ 85 %RH or less
Contact configuration
This card is a 2-wire 10 channel type for micro-current measurement, and is designed to keep the insulation resistance at 1014 Ω or higher. It is expandable up to 60 channels.
Specifications
Number of channels | 10 ch/2-wire |
Channel switch | Dry reed relay |
Peak voltage |
200 V (DC) |
Current | 1 A (DC) |
Insulating resistance (between terminals)* |
1014 Ω or higher (between HI terminals) |
Thermal electromotive force | 200 μV or less |
Number of required slots | 2 |
Applicable switch card | Integrated type |
*At 23±5 ℃ and 60 % RH or less
Contact configuration
This card consists of HI side relays (K1 to K10) and LO side relays (K11 to K20). The LO side relay operation can be selected from the following three modes.
LO-side relay operation mode | Sync Mode | Open Mode | Close Mode |
HI-side relay
Lo-side relay |
This card consists of HI side relays (K1 to K10) and LO side relays (K11 to K20). The LO side relay operation can be selected from the following three modes.
LO-side relay operation mode | Sync Mode |
HI-side relay
Lo-side relay
|
|
LO-side relay operation mode | Open Mode |
HI-side relay
Lo-side relay
|
|
LO-side relay operation mode | Close Mode |
HI-side relay
Lo-side relay
|
This card has switches of independent 4-wire 10 channels.
As the input and output terminal section and the switch section are integration, the 31002K can be uses as 2-wire 20 channel by setting.
Specifications
Number of channels |
10 ch/4-wire 20 ch/2-wire |
Channel switch | Mechanical relay |
Peak voltage |
250 V (AC) 110 V (DC) |
Current | 10 A |
Insulating resistance (between terminals)* |
1010 Ω or higher |
Thermal electromotive force | 50 μV or less |
Number of required slots | 2 |
Applicable switch card | Integrated type |
*At 23±5 ℃ and 60 % RH or less
Contact configuration
Current conducts between A and a, B and b, C and c, and D and d of each channel by setting ON (closing) arbitrary channels of CH0 to CH9.
The 31003B is a general-purpose switch card with 8×4-channel matrix configuration (4-wire) using mechanical relays. Tt can be used as 16×4-channel matrix configuration (2-wire) by setting.
Specifications
Number of channels | 8×4 ch/4-wire |
Channel switch | Mechanical relay |
Peak voltage |
125 V (DC) 110 V (AC) |
Current | 1 A |
Insulating resistance (between terminals)* |
1010 Ω or higher |
Thermal electromotive force | 50 μV or less |
Number of required slots | 1 |
Applicable switch card | 31009B |
*At 23±5 ℃ and 60 % RH or less
Contact configuration
Current conducts between arbitral channels of X (8 channels) and Y (4 channels) by setting ON (closing) the corresponding cross point between X-n and Y-n.
This card can be used as multiplexer card or actuator card depending on the selected terminal card. It is a 4-wire 10 channel type using mechanical relays for general purpose. It can be used as 2-wire 20 channel by setting.
Specifications
Number of channels |
10 ch/4-wire 20 ch/2-wire |
Channel switch | Mechanical relay |
Peak voltage |
125 V (DC) 110V (AC) |
Current | 1 A |
Insulating resistance (between terminals)* |
1010 Ω or higher |
Thermal electromotive force | 1 μV or less |
Number of required slots | 1 |
Applicable switch card | 31009B, 31009M |
*At 23±5 ℃ and 60 % RH or less
Contact configuration
This card can be used as contact configuration of the multiplexer card (31001 series) or the actuator card (31002 series) depending on the selected terminal card.
The 31009 is a terminal card for input and output signals that is used together with the 31001F, 31003B or 31005B in combination to be inseted in a scanner slot. It is a 4-wire type for general purpose.
Specifications
Input and output terminal | Screw type block terminal |
Number of terminals |
84 4×20 ch+4 (COM) |
Peak voltage betwen terminals |
200 V |
Maximum rated current | 1 A |
Insulation resistance between terminals* | 1010 Ω or higher |
Applicable switch card |
31001F, 31003B, 31005B |
*At 23±5 ℃ and 60 % RH or less
Contact configuration
The 31009 is a terminal card for input and output signals that is used together with the 31005B in combination to be inseted in a scanner slot.
Specifications
Input and output terminal | Screw type block terminal |
Number of terminals |
44 4×10 ch+4 (COM) |
Peak voltage betwen terminals | 200 V |
Maximum rated current | 1 A |
Insulation resistance between terminals* | 1010 Ω or higer |
Applicable switch card |
31005B |
*At 23±5 ℃ and 60 % RH or less
Contact configuration
Name | Part Number | Quantity |
Power cable (JIS 2 m) | A01402 | 1 |
ScanBus cabel (0.5 m): 3101 only |
CC012003-50 |
1 |
Name | Part Number |
Scan Bus cable (0.5 m) | CC012003-50 |
Scan Bus cable (2 m) | CC012003-200 |
Scan Bus cable (10 m) | CC012003-1000 |
Panel mount set (2U half) | A02039 |
Panel mount set (2U half twin) | A02040 |
Rack mount set (JIS 2U half) | A02263 |
Rack mount set (JIS 2U half twin) | A02264 |
Rack mount set (EIA 2U half) | A02463 |
Rack mount set (EIA 2U half twin) | A02464 |
Function: Multi-channel measurement using the digital multimeter 7461A/7461P, capable of meausirng DCV and 2-wire resistnace by sequential scan of up to 40 channels and 4-wire resistance by that of up to 20 channels.
*Go to the download page.